M3500B/M5500B Service Manual
3
-
7
Diagnosing Internal Failures
3
2. Select
Print Log
and press
.
3. The printer will print the System Log strip. See Figure 3-2.
4. Check the printout to verify the printed results are consistent with the
hardware in place. Verify options installed (SpO2, pacing) and the unit’s
serial number (on the bottom of the case). If the printout is not correct,
investigate and resolve the source of the mismatch. Then reset the hard-
ware options and serial number as needed using the Support Software
Tool. See "The Language Support Tool" on page 4-67.
5. If there are errors reported in the System Log:
a.
Check the time and date stamps to see if they are recent errors or not.
Consult Table 3-2 "Error Codes" on page 3-11 to identify the errors.
Then begin troubleshooting as needed (see the "Troubleshooting"
chapter.)
b.
Check that the time and date are correct. If they are not correct, access
Configuration Mode and reset them. See "Accessing Configuration
Mode" on page 3-5.
Figure 3-2 Sample System Log Printout
ENTER
SYSTEM LOG
Firmware Versions
Error Codes
Main:
34
90007 18:33 19 Jan 2000
DSP:
02
196:
57.00
Key:
06
SpO2:
02.42 01.04
Language:
English
Serial Number:
US0100241
Options:
Pacer SpO2
Shocks:
2
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...