WARNING
Do not immerse, or pour fluids on, any portion of the M3500B/M5500B.
WARNING
Do not use the M3500B/M5500B in a flammable or oxygen-rich atmosphere. This
can cause an explosion hazard.
WARNING
Avoid connecting the patient to several devices at once. Leakage current limits may
be exceeded. Do not use a second defibrillator on the patient while pacing with the
M3500B/M5500B.
WARNING
Avoid contact between the patient and metal objects, such as the gurney. Contact
with metal objects could cause unintentional current pathways.
WARNING
Operating the M3500B/M5500B or its accessories in conditions outside the environ-
mental specifications can result in device or accessory malfunction.
WARNING
Avoid exposing the display to direct sunlight on a hot day. Overheating can occur
and cause the display to black out, making the M3500B/M5500B temporarily unus-
able.
WARNING
Medical electrical equipment which does not incorporate defibrillator protection
should be disconnected during defibrillation.
WARNING
Electric shock hazards exist internally. Do not remove assembly screws. Refer ser-
vicing to qualified personnel.
CAUTION
This device has not been evaluated for use with electrosurgery equipment.
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...