User-replaceable Parts and Accessories
4
Carrying Case
Removal
To remove the carrying case from the M3500B/M5500B, follow steps 1
through 7 as shown in Figure 4-4. Note the screws are Phillips head, not Torx.
Also, after loosening the captive screws (step 6), lay the unit flat and lift the
defibrillator up off the mounting plates to disengage it from the 2 locating pins
in the carrying case. Then slide it out of the carrying case as shown in step 7.
Figure 4-4 Removing the Carrying Case