M3500B/M5500B Service Manual
7
-
5
Specifications
7
SYMBOLS:
Mark Triangle (for presses of the Mark key), an Alarm Bell
(Alarm Limit Violations), Lightening Bolt (Shock Delivered), Vertical stripe
Boundaries/Pacer/Sync Tick Marks).
Event Printing:
Mark Event key automatically documents events and ECG
during defibrillation episodes. The Mark Event key can annotate the event
with one of the following labels: Epinephrine, Atropine, Lidocaine, and
Other.
Auto Printing:
The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing:
The printer can be configured to run real time or with a six
second delay.
Reports:
The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log.
Speed:
25 mm/s with an accuracy of 5%.
Amplitude Accuracy:
10% or 50 uV.
Paper Size:
50 mm by 30 m (100 ft.).
Battery and Battery Power Modules
Battery Type
: 2 Ah, 12V, rechargeable, Sealed Lead Acid.
Battery Dimensions:
2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm (W) x 182 mm (D).
Weight:
1.4 lb. (0.65 kg).
Charge Time:
z
Approximately 14.5 hours to 100%, indicated by LED on power
module.
z
Approximately 2.5 hours to 90%, indicated by LED on power mod-
ule.
Capacity:
2.7 hours ECG monitoring or 50 full-energy discharges or 2.1
hours ECG monitoring while pacing (with a new, fully charged battery and
temperature above 20
o
C).
Battery Indicators:
LOW BATTERY message appears on monitor when at
least 10 minutes of monitoring time and 6 maximum-energy discharges
remain (with a new battery and temperature above 20
o
C).
Battery Storage:
Should not be stored above 40
o
C for extended periods of
time.
Charger Output:
Unit can be operated using Class I AC or DC power mod-
ule alone with no battery installed.
+
+
+
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...