Momentary Messages
3
-
18
Troubleshooting
Extended Self Test Failures
Should the unit report a
FAIL
in the Extended Self Test, resolve it using the
solutions below.
Table 3-6 Extended Self Test Failures
Failure
Possible Cause
Suggested Solution
RAM
ROM
System
CODEC
IRDA
Timebase
Failure on Control PCA.
Replace Control PCA.
System FAIL 10
Lithium backup battery failure.
Replace lithium battery.
Data Card
Data Card full, or incompatible, or
defective.
Replace Data Card with new, empty
M3510A/M5510A.
Failure on Control PCA.
Replace Control PCA.
Pacer
Defib
Failure on Power PCA.
Replace Power PCA.
Failure on Control PCA.
Replace Control PCA.
FE
Failure on Control PCA.
Replace Control PCA.
Failure on Power PCA.
Replace Power PCA.
Failure on Interface PCA.
Replace Interface PCA.
SpO2
Failure on SpO
2
PCA.
Replace SpO
2
PCA.
Failure on Control PCA.
Replace Control PCA.