Diagnostic Tests
2
-
30
Performance Verification and Safety Tests
Safety Tests
This section covers tests of the defibrillator’s electrical safety.
These safety tests only need to be performed if the customer normally uses the
defibrillator with the AC Power Module connected. If the defibrillator is nor-
mally used on battery power only, these tests do not need to be performed.
Using the procedures called out by the manufacturer of the analyzer in use,
measure and record the following data. Only test at the AC Mains (line) volt-
age used in the customer’s facility - there is no need to test both 120VAC and
240VAC. Also, test both Normal and Reverse Polarity line connections, and
record the worst case value.
Earth Leakage
z
Normal Condition - < 500 uA (< 300 uA for UL)
Record as "aaa".
z
Single Fault Condition - < 1000 uA
Record as "bbbb".
Patient Lead Leakage
Test both Leads (C) and pads (B) inputs.
z
Source (Normal Condition) - < 10 uA (C), < 100 uA (B)
Record as "ccc".
z
Source (Single Fault Condition) - < 50 uA (C), < 500 uA (B)
Record as "ddd".
z
With Mains on Applied Part (Single Fault Condition) - < 50 uA (C),
< 5000 uA (B)
Record as "eeee".
z
Auxiliary (Normal Condition) - < 10 uA (C), < 100 uA (B)
Record as "fff".
z
Auxiliary (Single Fault Condition) - < 50 uA (C), < 500 uA (B)
Record as "ggg".
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...