Error Codes
3
-
12
Troubleshooting
80000 - 8FFFF
IRDA failure.
1. Replace Interface PCA.
2. Replace Control PCA.
90000 - 90002
Self Test failure - RAM/ROM or Gate
Array.
1. Replace lithium battery.
2. Replace Control PCA.
90003
Self Test failure - Data Card circuits.
Replace Control PCA.
90004 and 90005
Self Test failure - Codec/time base.
1. Replace lithium battery.
2. Replace Control PCA.
90006
Self Test failure - SpO
2.
1. Replace SpO2 PCA.
2. Replace Control PCA.
3. Replace Interface PCA.
90007 and 90008
Self Test failure - Pacer or
Defib.This error can be caused by
pressing the Pacer key during the
self-test.
1. Run the self-test again, making
sure not to touch the unit during
the test.
2. Replace Power PCA.
3. Replace Control PCA.
90009
Self Test failure - Front End.
1. Replace Power PCA.
2. Replace Control PCA.
3. Replace Interface PCA.
9000A
Lithium backup battery failure.
1. Replace lithium battery.
2. Replace Control PCA.
A0000-A7FFF
Data Card failure.
1. Replace Data Card.
2. Replace Control PCA.
A8000 - AFFFF
Data Archival error.
1. Replace Data Card.
2. Replace Control PCA.
B0000 - BFFFF
Audio failure.
Replace Control PCA.
F0000 - FFFFF
RAM/ROM failure.
Replace Control PCA.
F0002
Promo Mode failure.
Replace Control PCA.
F0003
Keyscanner failure.
1. Replace Keyscan PCA.
2. Replace Control PCA.
F0004
System Communication error.
Replace Control PCA.
F0007
System Communication error
Cycle power.
Table 3-2 Error Codes (Continued)
Error Code
Meaning
Possible Solutions
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...