Performance Verification and Safety Tests
Overview
This chapter describes the tests and inspections required to verify perfor-
mance of the M3500B/M5500B Defibrillator/Monitor.
The information is presented in the following sequence:
Mandatory Testing
The Performance Verification Tests in this chapter are intended to verify
proper operation of the M3500B/M5500B following repair. The level of test-
ing required corresponds to the type of repair performed, and is divided into 3
categories:
z
External/No Trouble Found
z
Printer
z
Internal
External Repairs/No Trouble Found
External Repairs
are those involving the repair or replacement of one or
more of the items below.
No Trouble Found
applies when no malfunction
can be found, or when the problem appears to be due to improper use. In
either situation, the key point is that
the case has not been opened.
z
ECG cable
z
SpO
2
cable or sensor
z
Battery
Test Matrix
A list of the tests required to verify per-
formance of the M3500B/M5500B,
along with expected test results.
Test Equipment
A list of commercially available test, cal-
ibration and simulation devices needed
to perform performance verification
tests.
Performance Verification
Step-by-step instructions for performing
each of the Performance Verification
Tests.
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...