M3500B/M5500B Service Manual
2
-
7
Test and Inspection Matrix
2
Pads
If readings are as expected:
•
Source
(Normal Condition) - ggg
< 100 uA
•
Source (Single Fault
Condition) - hhh
< 500 uA
•
With Mains on applied part
(Single Fault condition) - iii
< 5000 uA
then Safety Pads Test
passes
Note: When recording test results, separate results within a test by slashes; separate tests by a semicolon (;); and
do not use empty spaces. For example:
VI:p;F:p;X:p;U:p;E:p;P:p;DM:p;DB:p;D:p;S1:p/200/300;S3:p/9/49/49/10/50/50/100/499/750
Table 2-1 Performance Verification and Safety Tests (Continued)
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
P (pass) or F (fail)