8 Troubleshooting
113
Battery Related Problems
Check Settings
INOP occurs during normal operation,
indicating a possible monitor software
problem
Check the monitor and patient settings
before you resume monitoring. If the
settings are unexpected, there may be a
problem with the monitor software.
1
Silence the INOP.
2
Load the User Defaults.
If this is unsuccessful, try loading the
Factory Default and reconfigure the
monitor in Configuration Mode, and save
the new settings in the User Defaults.
If the INOP persists, there is an unresolved
software problem. Report the problem to
factory support.
INOP occurs after a software upgrade,
indicating a possible incomplete or
unsuccessful upgrade
Clone the correct settings via the Support
Tool.
Check TI Config
INOP occurs after a software upgrade or
option change, indicating that the monitor
settings need to be updated to support the
software feature "NST Report"
Check the setting item
Setup NST Report
,
Guideline
. If set to "none", select the
appropriate NST Report Guideline. If no
Guideline is available, clone updated
settings via the Support Tool (e.g. settings
based on an initial configuration file for the
installed software revision, plus customer
specific adaptations, if required).
Internal.Comm.Malf
Main CPU Board defective
Replace the Main CPU Board.
Settings Malfunct
Problem during cloning process
Reclone the configuration file.
Memory space in which the settings are
stored has been corrupted
Reclone the configuration file. This will
reload the memory space.
Main CPU Board defective
Replace the Main CPU Board.
INOP Message
Possible Causes of Failure
Failure Isolation and Remedy
Symptoms
Causes of Failure
Failure Isolation and Remedy
Battery symbol is not displayed The monitor is not configured for
battery operation.
Make sure a system interface board with
battery capability is installed.
Batt Empty
INOP tone, battery
LED flashes
During this INOP, alarms
cannot be paused or switched
off
The estimated remaining battery-
powered operating time of the battery
is
≤
10 minutes.
Recharge the battery immediately.
If the condition persists, this INOP is re-
issued two minutes after you acknowledged
it.
Batt Incompat
INOP tone
The indicated battery cannot be used
with this monitor.
Replace with the correct battery (M4605A).
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...