7 Testing and Maintenance
88
Carrying Out and Reporting Tests
Test Report
Test and Inspection Matrix
Testing Organization:
Name of testing person:
(Check one of the following three options)
Test before putting into service (reference
value)
Recurrent Test
Test after Repair
Responsible Organization:
Device Under Test:
ID-Number
Product Number:
Serial No.:
Accessories:
Measurement Equipment (Manufacturer, Type,
Serial No.):
Functional Test (parameters tested):
Test Block
Test or Inspection
to be Performed
Expected Test Results
Record the Results (mandatory for
Philips Personnel only)
What to Record
Actual Results
Visual
Inspection
Perform Visual
Inspection -- see
“When to Perform
Tests” on page 56.
Pass or Fail
V:P or V:F
Power On
Power on the unit.
Does the self-test
complete
successfully?
If Yes, Power On test is passed.
PO:P or PO:F