7 Testing and Maintenance
95
4
Either:
attach electrodes to the M1363A adapter cable, and apply the electrodes to the skin (for example
on the wrists).
or
attach the M1363A adapter cable to a patient simulator.
Note
We do not recommend the use of a specific patient simulator. The use of a patient simulator does
not allow checking the specification of the ECG-functionality; it allows only a check of general
function.
Result: You should see MECG values displayed on the maternal display or annotated on the recorder
trace.
If the test results are not as outlined above, repeat the test with another ECG transducer. If this does
not solve the problem:
•
The MECG adapter cable may be defective. Replace the adapter cable, and repeat the test.
•
Check all connections.
Testing the Patient Module (M2738A)/Toco+ Transducer (M2735A): IUP Mode
1
To test the IUP functionality of the patient module or the Toco
+
transducer, you need the
following:
– Manometer.
– Expansion chamber.
– Three lengths of silicone tubing with a ‘T’ adapter.
2
Switch on the monitor and the recorder.
3
Connect the patient module or Toco
+
transducer to the fetal monitor.
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...