7 Testing and Maintenance
75
Case 1: Medical Device Combined with Medical Device
If you combine a medical device with another medical device (incl. Philips specified displays) to form a
medical electrical system according to IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010, no
additional protective measures are required. The medical electrical devices may be located in or outside
the patient vicinity in a medically used room. This is valid as long as the medical devices are connected
to separate mains outlets. No system test is required.
If the combined medical devices are connected to the same multiple portable socket outlet an
enclosure leakage current test of the entire device combination on the multiple portable socket outlet is
required to ensure that the resulting protective earth leakage current and equipment leakage current
does not exceed the limits of IEC 60601-1:2005+A1:2012/EN 60601-1:2006+AC:2010. Avoid using
multiple portable socket outlets. The medical electrical devices may be located in or outside the patient
vicinity in a medically used room. If the limits are exceeded, additional protective measures are
required, e.g. a separation device or the connection of each device to separate mains.
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...