12 Specifications
238
Recorder Specifications
Built-in Thermal Array Fetal Trace Recorder
Mechanism
Thermal Array Recorder
Paper & Printing
Type
Standard Z-fold paper
Standard Speeds (real-time traces)
3 cm/min, 2 cm/min, 1 cm/min
Fast Print Speed (stored traces)
Max. 20 mm/s
Print speed is variable and depends on
the print load
ECG Wave Print Speed
(not real-time)
Emulated 25 mm/s
Print speed is variable and depends on
the print load
Paper Advance
20 mm/s
Sensing
Optical Reflex Sensor for black page
marks
Accuracy @ 3 cm/min, 2 cm/min,
1 cm/min
±5 mm/page
Usable Print Width
128 mm
Resolution
8 dots/mm (200 dpi)
Time Delay to see trace on paper
<30s @ 1 cm/min
Trace Separation Offset for FHR
(Ultrasound and DECG)
Twin
Standard
FHR2 +20 bpm
Classic
FHR1 +20 bpm
Triplet
Standard
FHR2 +20 bpm
FHR3 -20 bpm
Classic
FHR1 +20 bpm
FHR3 -20 bpm
Recorder Symbols
Symbol
Description
Parameter is capable of alarming and alarms were enabled at the time of printing the annotation.
The low limit is printed before the symbol, and the high limit after it.
Parameter is capable of alarming, but alarms were disabled at the time of printing the annotation.
(Note: There is no alarm related annotation at all if a parameter does not have alarming
capability.)
FMP detection is on
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...