10 Repair and Disassembly
195
Removing the SpO2 Assembly
1
Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 141).
The SpO
2
assembly is identified in the next picture.
2
Remove the two screws holding the SpO
2
assembly.
3
Lift the side of the SpO
2
assembly nearest the SpO
2
socket, carefully disconnecting the multi-pin
connector shown.