7 Testing and Maintenance
98
Fetal Recorder Selftest Report
To verify your printer configuration, or if you doubt the performance of the recorder, you may want to
print a test report.
To print a selftest report, in Service Mode, select
Main Setup, Fetal Recorder, Selftest
.
Here is an excerpt from a sample test report to give you an idea what it looks like (the exact appearance
may vary slightly):
Check the test pattern to ensure all the heating elements on the printer head are operational. Ensure
that:
•
No more than 20 dots are missing over the entire printhead.
•
No more than two adjacent dots are inoperative.
•
No dots in the mode annotation (for example, FHR1) are inoperative.
1
Same data from the trace header
2
Self-test status
3
Print quality data
4
Recorder electrical data
5
Paper related data (paper count is the number of pages detected om this test).
6
Speed value is the tolerance for paper detection. If the Speed value is greater than 150, the
recorder most likely requires servicing (cleaning or exchanging of parts), assuming the correct
paper is being used. A paper value of 2984 is optimal. You may see small variants from this
number.
7
Expected value for PH Volt Min is 0. If the value is greater than 0 there may be a TLPH voltage
problem.
8
This line should print exactly on the 20 unit gridline (Toco baseline) if the recorder is correctly
calibrated.
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...