7 Testing and Maintenance
96
4
Attach the IUP adapter cable (989803143931) to the socket on the patient module or Toco
+
transducer.
5
Cut the sensor tip off an IUP catheter (M1333A).
6
Connect the catheter to the IUP adapter cable.
7
Connect the silicone tubing to the test volume chamber and the manometer as shown in the
picture.
8
Connect the cut end of the catheter to the silicone tubing.
9
Apply a pressure of 80 mmHg ± 5 mmHg with the manometer. Check that the value on the
display and on trace corresponds to this pressure. Slowly release the pressure, and check that the
value on the display and on trace shows this change in pressure.
Touchscreen Calibration
To access the touchscreen calibration screen:
1
Enter Service Mode
2
Select
Main Setup
3
Select
Hardware
4
Select
Calibrate Touch
Make sure you complete the calibration procedure without powering off the monitor mid-way. If the
monitor is powered off after the first point is touched, the touch panel will be deactivated until the
touch calibration is performed again.
If the touchscreen is accidentally mis-calibrated by selecting the wrong spot, you must use another
input device to re-enter calibration mode, for instance, a mouse connected to the PS/2 interface. If
you have the support tool, you can select
Start Calibrate Touch
and it will create a rough calibration
which will allow you to access the calibration menu again via the touchscreen.
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...