10 Repair and Disassembly
207
Exchanging the Noninvasive Blood Pressure/SpO2 Sockets
1
Remove the front end assembly -- see “Removing the Front End Assembly” on page 200.
2
Strip the front end assembly down as far as the end of step 6 on under “Removing the Front End
Assembly” on page 200, but without removing the fetal sensor socket connector block.
3
Take a new noninvasive blood pressure and/or SpO
2
socket as applicable. If you need to refit both
the noninvasive blood pressure and/or SpO
2
socket, you need to slot them together before
attaching them back onto the bus master board.
4
Reassemble the front end assembly -- see “Reassembling the Front End Assembly” on page 205.
Exchanging the Loudspeaker
1
Remove the top cover assembly (see “Removing the Top Cover Assembly” on page 141).
2
Disconnect the loudspeaker cable connector from the main CPU board.
3
Remove the three screws holding the loudspeaker, and remove the loudspeaker.
4
Refit the loudspeaker, making sure to refit the o-ring gasket. If a new gasket is required, it is
available as part of the bottom housing small parts kit. (See “FM Small Parts Kit - Screws and
Cables” on page 137).
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...