12 Specifications
233
Noninvasive Blood Pressure
Complies with IEC 80601-2-30:2009 / EN 80601-2-30:2010.
*1: Clinical investigation with the auscultatory reference method
•
The 5th Korotkoff sound (K5) was used to determine the diastolic reference pressures.
Response Time of Heart Rate
meter to Change in Heart Rate
M2735A, M2738A HR change from 80—120 bpm Average: 12 seconds
HR change from 80—40 bpm Average: 15 seconds
CL Toco
+
MP,
CL ECG/IUP
HR change from 80—120 bpm Average: 10 seconds
HR change from 80—40 bpm Average: 12 seconds
Heart Rate Meter Accuracy and
Response to Irregular Rhythm
M2735A, M2738A Ventricular bigeminy 40—60 bpm
Slow alternating ventricular bigeminy 45 bpm
Rapid alternating ventricular bigeminy 163 bpm
Bidirectional systoles 63—73 bpm
CL Toco
+
MP,
CL ECG/IUP
Ventricular bigeminy 40—60 bpm
Slow alternating ventricular bigeminy 30 bpm
Rapid alternating ventricular bigeminy 70—163 bpm
Bidirectional systoles 63—73 bpm
Maternal ECG Supplemental Information as required by IEC 60601-2-27
Performance Specifications
Measurement
Ranges
Systolic
30—270 mmHg (4—36 kPa)
Diastolic
10—245 mmHg (1.5—32 kPa)
Mean
20—255 mmHg (2.5—34 kPa)
Accuracy
1
Max. Std. Deviation:
8 mmHg (1.1 kPa)
Max. Mean Error: ±5 mmHg (±0.7 kPa)
Pulse Rate
Range
40—300 bpm
Accuracy
(average over
noninvasive blood
pressure measurement
cycle)
40—100 bpm: ±5 bpm
101—200 bpm: ±5% of reading
201—300 bpm: ±10% of reading
Measurement Time
Typical at HR >60 bpm
Auto/manual: 30 seconds (adult)
Maximum time: 180 seconds (adult)
Cuff Inflation Time
Typical for normal adult cuff: Less than 10 seconds
Initial Cuff Inflation Pressure
165 ±15 mmHg
Auto Mode Repetition Times
1, 2, 2.5, 3, 5, 10, 15, 20, 30, 45, 60, or 120 minutes
Venipuncture Mode Inflation
Inflation Pressure
20—120 mmHg (3—16 kPa)
Automatic deflation after
170 seconds
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...