2 Site Preparation
17
WARNING
NEVER run power cables through the same conduit or trunking used for system cables.
•
Installing wall boxes
It is the customer's responsibility to provide and install wall boxes to house face plates. The customer
must notify the Philips installation coordinator of which size is to be used.
Procedures for Philips Personnel
Before you begin the procedures in the installation sections, ensure that the customer has completed
all necessary preparations outlined in the previous section .
Site Requirements
The site requirements are listed in this section.
Space Requirements
The situating of the monitor should be planned such that the nursing staff are able to monitor the
patient with relative ease, with all patient connectors and controls readily available, and the displays
clearly visible. The location should also allow access to service personnel without excessive disruption,
and should have sufficient clearance all round to allow air circulation.
Dimensions and weight:
Monitor:
Size (W x H x D): 335 x 286 x 133 mm (13.2 x 11.3 x 5.2 in.)
Weight; 5.1 kg (11.2 lb.) with battery option #25 5.3 kg (11.7 lb.)
Transducer:
Size (diameter): 83 mm (3.27 in)
Weight (without cable): 190 g (6.7 oz)
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...