7 Testing and Maintenance
74
Required Protective Measures at System Installation
For any IT equipment (IEC 60950) operated in patient vicinity ensure that the equipment leakage
current does not exceed the limits described in IEC 60601-1. Use a separation device to ensure
compliance. After installation of IT equipment in patient vicinity, an enclosure leakage current test is
required.