8 Troubleshooting
114
Keyboard/Mouse Not Functioning
Network Status Icons
Batt Low
INOP tone
The estimated battery-powered
operating time remaining is less than
20 minutes.
Recharge the battery.
Battery Malfunct
INOP tone,
battery LED flashes
During this INOP, alarms
cannot be paused or switched
off, if the monitor is not
connected to AC power.
The monitor cannot determine the
battery status, or there is a
communication problem between the
battery and the battery board.
Replace the faulty battery. If the condition
persists, and the monitor is not connected
to AC power, this INOP is re-issued two
minutes after you acknowledged it.
Check the battery in a different monitor or
in a battery charger. If the INOP persists
the battery is faulty.
Check the battery board with known good
batteries. If the INOP persists, replace the
battery board.
If the problem persists, replace the main
board.
Charger Malfunct
INOP tone,
battery LED may flash
There is a problem with the battery
charger in the monitor.
Switch the monitor off and back on again. If
the problem persists replace battery with a
known good battery. If the INOP is shown
again, replace the battery board. If the
problem persists replace the main board.
Check Batt Temp
INOP tone
The temperature of the battery is too
high.
Check that monitor is not exposed to heat.
Symptoms
Causes of Failure
Failure Isolation and Remedy
Symptoms
Possible Causes of Failure
Failure Isolation and Remedy
Keyboard/Mouse
attached directly to the
monitor not functioning
Keyboard/Mouse not connected properly
Check cabling
Keyboard/Mouse defective
Replace Keyboard/Mouse
USB/PS/2 I/O board is not properly
plugged in
Ensure the USB/PS/2 I/O board is
properly plugged in. If necessary, remove
the board and plug it in again.
USB/PS/2 I/O board defective
Replace I/O board
Icon
Explanation
No Icon
LAN cable not connected (monitor does not have a LAN connection).
LAN cable connected, no connection to OB TraceVue/IntelliSpace Perinatal.
To check whether an IP address has been assigned, enter
Main Setup
,
Bed Information
and
scroll to
IP Address
.
0.0.0.0
means no IP address has been assigned.)
OB TraceVue/IntelliSpace Perinatal may send a prompt message, giving the reason why a
connection to OB TraceVue/IntelliSpace Perinatal cannot take place.
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...