7 Testing and Maintenance
69
Test to perform:
Measuring circuit for the measurement of Applied Part Leakage Current - Direct method
according to IEC/EN 62353
This test measures applied part leakage current from applied part to earth caused by external main
voltage on the applied part of 264 V. Each polarity combination possible shall be tested. This test is
applicable for ECG measurement inputs.
This safety test is based on IEC/EN 60601-1, IEC/EN 62353, and UL2601-1 Ed. 2/UL60601-
1:2003
.
For measurement limits and test voltage, refer to test block Safety (3), Option B71. Report the
highest value: (X1).
NOTE
All values for current and voltage are the root mean square (r.m.s.) values, unless otherwise stated.
Applied Part Leakage Current - Mains on Applied Part for Tympanic
Thermometer on Standalone Base Station
1
Remove the interface and the thermometer cable from the base station.
Test
Expected Test Results
Applied Part Leakage Current test (Single Fault
Condition - mains on applied part)
X1 <= 50
μ
A