12 Specifications
230
IUP
ECG
Complies with IEC 60601-2-27:2011 / EN 60601-2-27:2006+AC:2006 except clauses listed below:
•
201.6.2, 201.8.5.5
•
201.12.1.101
•
202.6.2.101
Performance Specifications
Maternal Pulse from Toco
Emitted Light Energy
≤
15 mW
Wavelength Range
780—1100 nm
Range
40—240 bpm
Resolution
1 bpm
Display Update Rate
1 per second
Accuracy
± 2% or 1 bpm, whichever is greater
Update Rate
every 4 seconds
Performance Specifications
IUP
Measurement Method
Passive Resistive Strain Gauge Elements
Measurement Range
-100—+300 mmHg
Signal Range
-99—127 mmHg or (-13.2—16.9 kPa)
Resolution
Display
1 mmHg
Printer
1/4 mmHg
Sensitivity
5
μ
V/V/mmHg
Offset Compensation
+100—-200 mmHg
Accuracy (not including sensor accuracy)
±0.5% per 100 mmHg
Update Rate
Display
1 per second
Printer
~4 per second
Auto Offset Correction
3 seconds after connecting the transducer, the IUP value is
set to 0 mmHg
Performance Specifications
ECG
Performance Specifications Type
DECG
Single Lead ECG (derived from Fetal Scalp
Electrode)
MECG
Single Lead ECG (derived from RA and LA
electrodes)
Measurement Range
30—240 bpm
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...