2 Site Preparation
18
Environmental Requirements
The environment where the FM20/FM30 monitor will be used should be reasonably free from
vibration, dust, and corrosive or explosive gases. The ambient operating and storage conditions for the
FM20/FM30 monitor must be observed. If these conditions are not met, the accuracy of the system
will be affected and damage can occur.
Monitor (M2702A/M2703A); Interface Cable for Avalon CTS (M2731-60001)
Temperature Range
Operating
without battery:
0°C — 45°C (32°F — 113°F)
with fully charged battery:
0°C — 40°C (32°F — 104°F)
with battery charging:
0°C — 35°C (32°F — 95°F)
Storage
-20°C — 60°C (-4°F — 140°F)
Humidity Range
Operating
<95% relative humidity @ 40°C (104°F)
Storage
<90% relative humidity @ 60°C (140°F)
Altitude Range
Operating
-500 — 3.000 m (-1.640 — 9.840 ft.)
Storage
-500 — 13100 m (-1.640 to 43.000 ft.)
Transducers (M2734A/M2734B/M2735A/M2736A/M2738A)
Temperature Range
Operating
0°C — 40°C (32°F — 104°F)
Storage
-20°C — 60°C (-4°F — 140°F)
Humidity Range
Operating
<95% relative humidity @ 40°C (104°F)
Storage
<90% relative humidity @ 60°C (140°F)
Altitude Range
Operating
-500 — 3.000 m (-1.640 — 9.840 ft.)
Storage
-500 — 13.100 m (-1.640 — 43.000 ft.)
SpO
2
Sensors
Operating Temperature Range
0°C — 37°C (32°F — 98.6°F)
Summary of Contents for Avalon FM20
Page 2: ......
Page 6: ...6 ...
Page 38: ...3 Installation Instructions 38 ...
Page 46: ...4 Theory of Operation 46 ...
Page 155: ...10 Repair and Disassembly 155 11 Put the two halves of the display assembly together as shown ...
Page 167: ...10 Repair and Disassembly 167 10 Remove the four screws holding the recorder chassis ...
Page 169: ...10 Repair and Disassembly 169 12 Remove the RFI bracket 13 Lift out the recorder chassis ...
Page 254: ...12 Specifications 254 ...
Page 259: ......
Page 260: ...Part Number 453564541271 Published in Germany 11 14 453564541271 ...