11 Upgrades
220
9
Fit the new connector cover that comes with the upgrade kit as described in step 8 for Option B71
on
10
Replace the reassembled front end assembly as described in “Setting the Fetal Recorder Offset” on
page 97.
11
Fit the SpO
2
board holder to the bottom housing with two screws from the upgrade kit.
12
Next, fit the SpO
2
board as described in “Replacing the SpO2 Assembly” on page 196.
13
Then refit the noninvasive blood pressure assembly as described in “Replacing the Noninvasive
Blood Pressure Assembly” on page 192.
14
Replace the top cover assembly as described in “Replacing the Top Cover Assembly” on page 180.
15
For tests to perform after upgrading, see “When to Perform Tests” on page 56.