Control interfaces
57
8.6.2
Wiring of digital inputs
Fig. 19: J5 connector: digital input wiring
8.6.3
Wiring of analog inputs
Fig. 20: J5 connector: wiring of analog inputs
8.6.4
Wiring of the digital outputs
Contact
Function
S1 (1-4)
Roots 2 frequency
converter status
Contact closed: Roots 2 starts.
Contact open: Roots 2 stops.
S2 (2-4)
Roots 2 frequency
converter warning
Contact closed: no warning
Contact open: there is a warning on Roots 2.
S3 (3-4)
not used
1
2
3
4
5
6
7
8
15
14
13
12
9
10
11
S2
GND
S1
S3
GND
S5
Contact
Function
A1 (10-11) Roots 2 temperature
Used to connect a temperature probe and to monitor it, setting
warning and alarm thresholds and a triggering timer, see parameter
setting
[SETTINGS] [TEMPERATURE] [ROOT 2 T°]
A2 (12-13) Roots 2 power
Setpoint between 0 and 10 V.
Used to monitor Roots 2 power by setting warning and alarm thresh-
olds and a triggering timer: see parameter setting
[SETTINGS]
[POWER] [R2 POWER]
RX /TX
(14-15)
RS485 Reception/
Transmission data
RS485 communication.
1
2
3
4
5
6
7
8
15
14
13
12
9
10
11
GND
TX
RX
A1
A2
GND
Contact
Function
7-9
Roots 2 start-up (contact closed whent Roots starts)
8-9
IDLE mode management
– contact open: maximum speed of rotation
– contact closed: reduced speed of rotation
5-6 (S5)
Roots 2 frequency converter alarm
Contact closed: no alarm.
Contact open: there is an alarm on Roots 2.
If Roots 2 alarm musn’t stop the pump, contact 5-6 must be strapped.