66
Malfunctions
11
Malfunctions
Study the safety instructions related to maintenance
11.1 What happens in the event of a defect
When a problem occurs, the user is informed through:
– the relevant fault indicator light: yellow lit in case of warning, red lit in case of alarm,
– the audible buzzer warning (if preselected),
– the fault contacts on
J1
remote connector are activated,
– display of fault message on the HHR,
– a message via the RS-232 or RS-485 serial link.
Messages are listed in the
Malfunctions
chapter
11.1.1 Meaning of light indicators
Indicator
light
Name
Status
Pump status
White
Power
Off
The pump is not powered
Lit
The pump is powered
Green
Running
Off
The pump is not running (Pump shut-down)
Lit
The pump is running
Yellow
Warning
Off
No warning
Blinking
Pre-warning.
Read the fault message on HHR.
Reset the fault pushing
ENTER
key.
Flashing
It was a pre-warning. .
Read the fault message on HHR.
Remedy to the trouble to delete the fault.
Lit
Warning presence.
Read the fault message on HHR.
Remedy to the trouble to delete the fault.
Red
Alarm
Off
No alarm
Blinking
There is a pre-alarm, pump is running.
Read the fault message on HHR.
Reset the fault pushing
ENTER
key.
Flashing
It was a pre-alarm.
Read the fault message on HHR.
Remedy to the trouble to delete the fault.
Lit
Alarm presence. The pump is stopped.
Read the fault message on HHR.
Remedy to the problem to delete the defect.
Switch power off and restart the pump
.