48
Control interfaces
SETTING
(1)
Selection
Setting range
CONTRAST
30-250
PURGE
PURGE FLOW
LOW DEFAULTS
PRE-WARNING DISABLED
ENABLED
Threshold
20-120 slm
Delay
0- 600 s
WARNING
DISABLED
ENABLED
Threshold
15-120 slm
Delay
0- 600 s
PRE-ALARM
Threshold
15 slm
Delay
0- 600 s
PURGE PROLONG
Delay
0-7200 s
WATER FLOW
SENSOR TYPE/
ANA. SENSOR
LOW DEFAULTS
WARNING
DESABLED
ENABLED
Threshold
130 to 960 l/h
Delay
0-600 s
DESABLED
POWER
FB POWER
HIGH DEFAULTS ALARM
DISABLED
ENABLED
Threshold
4000-20000 W
Delay
0-6500 s
WARNING
DISABLED
ENABLED
Threshold
500-4900W
Delay
0-600 s
PRE-WARNING DISABLED
ENABLED
Threshold
500-4000W
Delay
0-600 s
R1 POWER
ALARM
DISABLED
ENABLED
Threshold
3000-20000 W
Delay
0-600 s
WARNING
DISABLED
ENABLED
Threshold
500-4000W
Delay
0-600 s
PRE-WARNING DISABLED
ENABLED
Threshold
500-3000W
Delay
0-600 s
R2 POWER
(2)
ALARM
DISABLED
ENABLED
Threshold
3000-20000 W
Delay
0-600 s
WARNING
DISABLED
ENABLED
Threshold
500-4000W
Delay
0-600 s
PRE-WARNING DISABLED
ENABLED
Threshold
500-3000W
Delay
0-600 s
FREQUENCY
FB INVERTER
MODE
FIXE
ANALOG
FIXE FREQ.
Min/Inter./Max
ANALOG FREQ.
R1 INVERTER
MODE
FIXE
ANALOG
FIXE FREQ.
Min/Inter./Max
ANALOG FREQ.
MODE
FIXE
ANALOG
R2 INVERTER
(2)
FIXE FREQ.
High /Low
ANALOG FREQ.
TEMPERATURE
HP FB T°
DISABLED
CONTROL
SETPOINT
H: 70-130 °C
X: 50-130 °C
DEFAULTS
HIGH DEFAULTS ALARM
Threshold
80-140 °C
Delay
0-600 s
WARNING
Threshold
se5 °C
Delay
0 s
LOW DEFAULTS DISABLED
WARNING
ENABLED
Threshold
setpoint - 10 °C
Delay
0 s
MEASUREMENT DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
ENABLED
Threshold
H: 75-140 °C
X: 55-140 °C
Delay
0-600 s
WARNING
DISABLED
ENABLED
Threshold
H: 75-140 °C
X: 55-140 °C
Delay
0-600 s