Advanced settings
7.11 Analog input management
An analog input (0-10 Volt)
[ANA1]
is available on the ANA1 plug
.
An analog input (0-10 Volt)
[USER ANA]
is available on the
J3
connector
2 monitoring thresholds (warning and alarm) and a delay linked to each threshold can be
set
[SETTING] [ANALOGIC INPUT] [ANA1]/[USER ANA] [ENABLED]
A temperature probe PT100 can be connected on
J3
. 2 monitoring
thresholds (warning and alarm) and a delay linked to each threshold can be set
[SET-
TING] [TEMPERATURE] [USER T°]
.
If the measured value exceeds the set threshold, a defect (warning or alarm) appears
after the expiry of a delay. An alarm message leads to pump shutdown.
If
[SETTING] [ANALOGIC INPUT] [ANA1]/[USER ANA] [DISABLED]
is set, there is no
threshold to configure and the fault is not managed.
7.12 Logic input management
Three logic inputs are available on the LI1, LI2 and LI3 plugs
LI1, LI2 and LI3 can be set opening or closing logic input
[SENSOR TYPE] [NO]/[NC]
.
Two monitoring delays can be set (warning and alarm)
[SETTING] [LOGIC INPUT]
[LOG1]/[LOG2]/[LOG3] [ENABLED]
.
When the contact LI1, LI2 or LI3 opens or closes, a fault (warning or alarm) appears after
the expiry of a delay.
If the fault is set to
[DISABLED]
, the fault is not managed.
Analog input (V)
Time
T. warning
Alarm
threshold
Warning
threshold
No message
T. alarm
Warning message
Alarm m
pump stopped
alarm
Logic input (LI1/LI2/LI3)
Time
warning
[Sensor type][NC]
1
0
No message
Warning
message
Alarm message
+ pump stopped
alarm
Logic input (LI1/LI2/LI3)
Time
warning
[Sensor type][NO]
1
0
No message
Warning
message
Alarm message
+ pump stopped