Operation
6.2.2
Pumping start-up
Press
START
key, valid with
ENTER
, the pumping starts. The
Running
green indica-
tor lights on.
The following operations will perform automatically:
– FB pump starts up,
– purge gas valve opens,
– Roots pump starts up,
– inlet valve opens (if any and if set).
The operating parameters are displayed.
Gain access to other operating parameters using
+
or
–
keys.
Only installed and set functions are displayed.
6.2.3
Auto-restart configuration
By default setting, automatic restart is set to
[DISABLED]
6.2.4
Start-up after an emergency stop
To restart the pump after an emergency stop, it is necessary to:
fix the problem,
rotate the emergency button
to unlock,
position the main switch
to
O
than to
I
position,
press
START
to start the pump and valide with
ENTER
.
6.2.5
Option Roots command
To start or stop the Roots while the FB is running, set the
[CMD.ROOTS]
on
[ENABLED]
when the pump is stopped.
The Roots will start or stop by pressing the
ROOTS
key if
[CONTROL MODE]
is
[HHR]
J1-S3
contact if
[CONTROL MODE]
is
.
6.3
Pump shut-down
6.3.1
Pumping stop
Press on
STOP
key and valid with
ENTER
.
The following operations will perform automatically:
– inlet valve closes (if any and if set),
– pump(s) stop(s) (FB and Roots),
– injection of purge gas continues during the shutdown phase if timing of
[PURGE PRO-
LONG]
.
10h
RUNNING FIXE SPEED
BF FREQ: 60 Hz
R1 FREQ: 100 Hz
HHR
R P
DANGER
Risk due to auto-restart
When the pump has been set for automatic restart after a power failure, the pump re-
starts when the power is switched back on.
Take all the measures required to prevent risks resulting from this type of operation.