76
Technical data and dimensions
(1)
According to CE regulations. Pumps can withstand a supply voltage variation of ± 10 %.
(2)
Measured at 1 m length. and 1.6 m height. according to ISO 9614-2 standard (pump equipped with
a silencer and a check valve).
(3)
This pump is certified for pumping a volume of 1 m
3
, with 20-minute cycling. In case of larger vol-
ume or a higher cycling frequency, consult us.
(4)
Subject to pump temperature setting.
(5) At maximum rotational speed.
(6)
Without inlet pipe extension, inlet flange is ISO-F type.
(7)
With standard N2 injection device. In case of specific N2 injection device, consult us.
Characteristics
Unit
A 2504H/A 2504X
A 3004H/A 3004X
Peak pumping speed (60 Hz)
m
3
/h
2500
3000
Maximum ultimate pressure
(without N
2
purge) (60 Hz)
hPa
4·10
-3
4 ·10
-3
Maximum ultimate pressure
(with N
2
purge 50 slm)
(7)
(60 Hz)
hPa
9·10
-3
9 ·10
-3
Maximum pumped volume
(3)
m
3
2
2
Maximum continuous inlet flow
(5)
50Hz /
60Hz
(5)
EHT version (DSB 3000 hot)
ELT version (DSB 3000 cold)
slm
90
100
80
100
Supply voltage
(1)
V
200-230 V / 380-480 V - 3 phases - 50/60 Hz
Power consumption at ultimate pressure
(50/60 Hz) (no exhaust heater)
kW
2,4/2,6
2,4 /2,6
Power consumption at max. inlet flow
kW
8
8
Full load current (200-230 V)
A
32
32
Full load current (380-480 V)
A
21
21
Maximum exhaust pressure
hPa
1200
1200
Cooling water flow
(4)
l/h
150-300
150-300
N
2
purge flow
slm
10 à 120
10 à 120
Oil fluid capacity
l
BF : 0,28
Roots : 1,5
BF : 028
Roots : 1,5
Helium localized leak rate
hPa l/s
< 1 ·10
-5
< 1 ·10
-5
Inlet flange
(6)
DN 160 ISO-K
DN 160 ISO-K
Exhaust flange
DN 40 ISO-KF
DN 40 ISO-KF
Dimensions
Weight
kg
800
800
Typical vibration level at inlet flange
g
< 0,1
< 0,1
mm/s
< 1,5
< 1,5
Maximum sound level
(2)
dB(A)
< 72
< 72
Summary of Contents for A 1204H
Page 88: ...ETL Mark...
Page 89: ...Certificate of conformance to SEMI...