28
Installation
Before switching on, connect the cover plug on EMS connector
To integrate the pump into the emergency stop circuit of a equipment, wired the EMS
connector at the rear of the pump
Semi S2 requirements (EMO)
To meet the requirements of the Semi S2 standards, it is the customer's responsibility to
integrate the pump into the equipment's emergency off circuit.
Use the EMS contact of the pump to control the equipment power failure (EMO)
5.9
Operating for the first time
Check before each time the pump is started up that the oil levels in the housings are
visible in the center of the sight glasses
: carry out this check with the pump
stopped.
Check that the connections to the pumping line, inlet and exhaust sides, have been
made
.
5.9.1
Switch on
Connect the hand-held remote module
HHR
.
Position the main switch
to
I
position. The white
Pow-
er
indicator lights on, the HHR displays the status of the
pump
Take the pump control via the
CONT.
key: HHR flashes
on the display.
5.9.2
Check of the rotation’s direction at initial start-up
Start-up the pump by pressing
START
and valid with
ENTER
:
– the pump starts, PUMPING is displayed on the HHR. Stop the pump pressing
STOP
and valid with
ENTER.
– the pump doesn’t start and the message D19 REVERSE UVW is displayed : there
is a wiring problem. Stop the pump pressing
STOP
. Switch off the pump by posi-
tioning the main switch to
O
: the white indicator light
Power
turns off. Invert two
phases at the main power input connector
185h
PUMPING STOPPED
05/09/16 10:10
HHR
R P
WARNING
Electric shock hazard in case of contact
When the main isolator is set at
O,
items located between the mains connection and the
isolator are still under mains voltage.
Disconnect the power line cable from all power sources before doing any work on the
product.
WARNING
Electric shock hazard
Some components have capacitors charged to over 60VDC. When power is switched
off, they maintain this charge for some time. Residual voltages from the filter capacitors
can cause electric shocks all the way back to the mains plug.
Wait 5 minutes after power-off before commencing any work on the product.