Control interfaces
49
TEMPERATURE
LP FB T°
ENABLED
CONTROL
SETPOINT
H: 70-120 °C
X: 50-120 °C
DEFAULTS
HIGH DEFAULTS ALARM
Threshold
80-130 °C
Delay
0-600 s
WARNING
Threshold
se5 °C
Delay
0 s
LOW DEFAULTS DESABLED
ENABLED
Threshold
setpoint - 10 °C
Delay
0 s
MEASUREMENT DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
ENABLED
Threshold
H: 75-130 °C
X: 55-130 °C
Delay
0-600 s
WARNING
DISABLED
ENABLED
Threshold
H: 75-130 °C
X: 55-130 °C
Delay
0-600 s
TEMPERATURE
ROOTS 1 T°
DISABLED
CONTROL
SETPOINT
50-100 °C
DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
ENABLED
Threshold
60-110 °C
Delay
0-600 s
PRE-ALARM DISABLED
ENABLED
Threshold
55-110 °C
Delay
0-600 s
WARNING
Threshold
se5 °C
Delay
0 s
LOW DEFAULTS WARNING
DISABLED
ENABLED
Threshold
setpoint - 10 °C
Delay
0s
MEASUREMENT DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
ENABLED
Threshold
60-110 °C
Delay
0-600 s
PRE-ALARM DISABLED
ENABLED
Threshold
55-110 °C
Delay
0-600 s
WARNING
DISABLED
ENABLED
Threshold
55-110 °C
Delay
0-600 s
VALVE STATUS
OPEN VALVE
CLOSE VALVE
TEMPERATURE
ROOTS 2 T°
(2)
DISABLED
MEASUREMENT DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
Threshold
110-140 °C
Delay
0-600 s
WARNING
DISABLED
Threshold
70-140 °C
Delay
0 s
HEATER 1 T°
DISABLED
CONTROL
SETPOINT
80-140 °C
DEFAULTS
HIGH DEFAULTS WARNING
Threshold
90-150 °C
Delay
0-600 s
LOW DEFAULTS WARNING
Threshold
setpoint - 10 °C
Delay
0 s
HEATER 2 T°
DISABLED
CONTROL
SETPOINT
80-140 °C
DEFAULTS
HIGH DEFAULTS WARNING
Threshold
90-150 °C
Delay
0-600 s
LOW DEFAULTS WARNING
Threshold
setpoint - 10 °C
Delay
0 s
HEATER 3 T°
DISABLED
CONTROL
SETPOINT
80-140 °C
DEFAULTS
HIGH DEFAULTS WARNING
Threshold
90-150 °C
Delay
0-600 s
LOW DEFAULTS WARNING
Threshold
setpoint - 10 °C
Delay
0 s
USER T°
DISABLED
MEASUREMENT DEFAULTS
HIGH DEFAULTS ALARM
DISABLED
Threshold
0-200 °C
Delay
0-600 s
WARNING
DISABLED
Threshold
0-150 °C
Delay
0-600 s
SETTING
(1)
Selection
Setting range