56
Control interfaces
8.5.2
Wiring of digital inputs
Fig. 17: J4 connector: digital input wiring
8.5.3
Wiring of the digital outputs
Fig. 18: J4 connector: valve wiring
8.6
Remote control J5 connector
8.6.1
Description
Connection made via the connector
J5
(Male D-Sub -15 pin connector)
– remote control of external Roots (Roots 2) pumping functions, start, stop and IDLE
mode,
– input of the external Roots speed setpoint, temperature and its monitoring.
1
2
3
4
5
6
7
8
15
14
13
12
9
10
11
S1
+24 VDC
GND
GND
S2
S6
+24 VDC
S7
Contact
Function
S1 (2-10)
IN_IN_VALVE_OPEN
Contact open in end of travel not reached: warning triggered.
S2 (3-11)
IN_IN_VALVE_CLOSE
Contact open in end of travel not reached: warning triggered.
S6 (6-14)
IN_SOFT_VALVE_OPEN
Contact open in end of travel not reached: warning triggered.
S7 (7-15)
IN_SOFT_VALVE_CLOSE
Contact open in end of travel not reached: warning triggered.
1
2
3
4
5
6
7
8
15
14
13
12
9
10
11
+
24
V
D
C
V1
+
24
V
D
C
V2
+ 5 VA
+ 5 VA
Contact
Function
1-9
OUT_INLET_VALVE
Inlet valve
V1
connection. Valve opens when the pump is switching on.
5-13
OUT_SOFT_VALVE
“Soft pumping” valve
V2
connection. Valve opens when the pump is switching on.
NOTICE
Extra low voltage circuit
The remote control circuits are equipped with dry contact outputs (30 V - 1 A max.).
Any overvoltage or overcurrent can result in internal electrical damage. The installer
must:
Connect these outputs in accordance with the rules and protection of safety extra low
voltage (SELV) circuits,
Supply these contacts with voltage of less than 30 V and current of less than 1 A.
Summary of Contents for A 1204H