Malfunctions
69
or
HP T° < setpoint T -10°C
HP FB water valve problem
Check the condition of the solenoid valve dia-
phragm and its operation.
Temperature sensor problem
Check the proper operation and the connection of
the sensor on FB HP TEMP plug
.
Other problem
Contact your service center.
W17/D17 T° R1 MOT
Roots 1 motor temperature is > 120°C
Check the water cooling supply on the pump.
Check that the Roots 1 motor is correctly cooled.
Check that the cooling pipes are not clogged.
Sensor or wiring faulty
Check the connection of the sensor on
RSV1 MOTOR plug
and in the motor
terminal box.
Problem on Roots 1 motor temperature sensors
Check the proper operation of Roots 1 motor tem-
perature sensors (NC contacts).
Other problem
Contact your service center.
D19 REVERSED UVW
Main power supply problem: two inverted phases
Check the wiring at the main power input connec-
tor.
Other problem
Contact your service center.
W21 SOFT P. VALVE
Incorrect monitoring configuration
Check that [DEFINITION] [INLET VALVE] is set on
[DESABLED] when no valve is connected.
Wrong connection
Check the valve connection on the
J4
connector
.
Malfunction
Check the proper operation opening/closing of the
valve.
Other problem
Contact your service center.
W31 SUPPLY 15V
15V supply problem
Contact your service center.
W32 MAINTENANCE
Incorrect monitoring configuration
Check the maintenance thresholds.
The maintenance threshold is reached
Contact your service center for pump overhaul.
D33 SUPPLY 24VS
24V supply problem
Check that the EMS is not switched on.
Check the EMS wiring circuit.
Check the 24 V power supply and the 24 V
PCboard.
D34 EMO
Pump emergency stop is engaged
Rotate to unlock the emergency stop on the front
panel display.
Emergency cover plug is missing
Check that the cover plug is fitted on the emergen-
cy connector at the rear of the pump.
Problem with 24 V power supply
Check the EMO power circuit.
Other problem
Contact your service center.
D36 FB CONTACTOR
Incorrect monitoring configuration
Check that [DEFINITION] [FREQUENCY IN-
VERTER] [FB INVERTER] is set on [DISABLED].
KM1 contactor not closed
Check that there is a strap between pins 2 and 3
of emergengy stop terminal
EMS
.
The FB does not start: check the 24 V power sup-
ply on A1 and A2.
Other problem
Contact your service center.
D37 R1 BREAKER
Incorrect monitoring configuration
Check that the [DEFINITION] [PUMP MODEL] is
set on [A204] or [A124].
Change the setting by sending an order via Mod-
bus communication protocole
(see
Serial link
op-
erating instructions).
RT2 (Roots) breaker switched off
Check RT2 setting
. Switch RT2 on.
Check that the presence of current and voltage on
the phases.
Roots 1 mechanical problem
Check the rotation of the Roots 1 rotors (if possi-
ble).
Other problem
Contact your service center.
D38 FB BREAKER
Incorrect monitoring configuration
Check that the [DEFINITION] [FREQUENCY IN-
VERTER][R1 INVERTER] is set on [DESABLED].
RT1 (BF) breaker switched off
Check RT1
. Switch RT1 on.
Check that the presence of current and voltage on
the phases.
FB mechanical problem
Free the FB manually
.
Other problem
Contact your service center.
D39 R1 CONTACTOR
Incorrect monitoring configuration
Check that the [DEFINITION] [FREQUENCY IN-
VERTER][R1 INVERTER] is set on [DESABLED].
KM2 (Roots) contactor not closed
The Roots 1 does not start: check the 24 V power
supply on A1 and A2.
Other problem
Contact your service center.
Incident
Cause
Remedy