54
Control interfaces
8.4
Remote control J3 connector
8.4.1
Description
Connection made via the connector
J3
(Male D-Sub - 15 pin connector)
allows:
– Remote control of pump functions: start, stop and IDLE mode,
– input of the speed setpoint and its monitoring,
– input of a temperature probe and monitoring of this temperature.
8.4.2
Wiring of digital inputs
Fig. 15: J3 connector : digital input wiring
NOTICE
Extra low voltage circuit
The remote control circuits are equipped with dry contact outputs (30 V - 1 A max.).
Any overvoltage or overcurrent can result in internal electrical damage. The installer
must:
Connect these outputs in accordance with the rules and protection of safety extra low
voltage (SELV) circuits,
Supply these contacts with voltage of less than 30 V and current of less than 1 A.
Contact
Function
S1 (1-9)
USER1
[DEFINITION] [ IDLE MODE] [ENABLED]
Selection mode IDLE, see table below.
S2 (2-10)
USER2
S3 (3-11)
USER3
S4 (7-8)
Pump Start/Stop
Contact closed: pump starts.
Contact open: pump stops.
This function is valid when
[DEFINITION] [CONTROL MODE] [US-
ER]
is selected
.
IDLE option
Contact status
Stand-by valve
Rotation speed
HHR display
S1
S2
S3
Full
1
0
0
closed
minimum
IDLE MODE
Intermediate
0
1
0
closed
intermediate
STD BY N2
Reduce speed
0
0
1
opened
minimum
IDLE MODE
Stand-by purge
1
1
1
closed
maximum
STD BY N2
Standard
0
0
0
opened
maximum
--
1
2
3
4
5
6
7
8
15
14
13
12
9
10
11
S1
S2
+24V
GND
S3
S4