46
Control interfaces
8
Control interfaces
This chapter describes the connections and protocols associated with each control
mode.
8.1
Different control modes
4 control mode are available:
– HHR:
locally
controlled by the hand-held remote,
– REMOTE:
remote controlled
by
J1
connector,
– SERIAL LINK:
remote controlled
by serial link
J2
,
J7
,
J6
,
– USER:
remote controlled
by
J3
connector.
HHR
The pump is controlled with the hand held remote connected in front
HHR
or at the rear
panel of the pump
J8
. For safety reasons, the pump can be controlled only by one HHR
at the same time.
REMOTE
The pump is remote-controlled through the opening or closing of different dry contacts
wired on remote control connector
J1
.
SERIAL LINK
The pump is remote-controlled by the commands transmitted via the serial link (see
Se-
rial link
Operating instructions).
USER
The pump is remote-controlled through the opening or closing of different dry contacts
wired on remote control connector
J3
.
The control mode is selected with HHR
[DEFINITION] [CONTROL MODE]
:
To make the pump running,
it is essential to install the cover plug delivered with the
pump to the emergency connector
or connect the pump to the emergency stop circuit
8.2
HHR control
When the pump is delivered, it runs according to the default setting: it depends on the
product ordering guide.
Menu
Description
DEFINITION
To set the pump model, functional settings, and external sensor management.
SETTING
To adjust the monitoring parameters.
MAINTENANCE
To read or adjust the pump maintenance time.
MANAGEMENT
To change the password, register the pump serial number and software versions.
LAST EVENTS
To display the events recorded.