86
Appendix
15.3 Circuit breaker setting
Fig. 29: Location of electrical components: pump with one frequency converter
VAR1
Frequency converter
A1
Current sensor
F1
Power filter
RT1
FB circuit breaker
CA
Auxiliary contact
RT2
Roots circuit breaker
KM1
FB command
RT3
Heating circuit breaker (option)
Fig. 30: Location of electrical components: pump with two frequency converters
VAR1/VAR2
Frequency converter
RT1
FB circuit breaker
F1/F2
Power filter
RT2
Roots circuit breaker
RT3
Heating circuit breaker (option)
VAR1
RT3
F1
RT2 RT1
CA
KM1
A1
RT3
RT2 RT1
VAR1
VAR2
F1
F2
Pump
Voltage
RT1
RT2
RT3
A 124H
LV
200-230 V - 50/60 Hz
14 A
-
4 A
HV
380-480 V - 50/60 Hz
9 A
-
2.5 A
A 204H/A 204X
LV
200-230 V - 50/60 Hz
16 A
-
4 A
HV
380-480 V - 50/60 Hz
13 A
-
2.5 A
A 604H
LV
200-230 V - 50/60 Hz
14 A
10 A
4 A
HV
380-480 V - 50/60 Hz
9 A
6 A
2.5 A
A 804H/A 804X
LV
200-230 V - 50/60 Hz
16 A
10 A
4 A
HV
380-480 V - 50/60 Hz
13 A
6 A
2.5 A
A 1204H
LV
200-230 V - 50/60 Hz
14 A
18 A
4 A
HV
380-480 V - 50/60 Hz
9 A
13 A
2.5 A
A 1504H/A 1504X/A 1804H/A 1804X
LV
200-230 V - 50/60 Hz
16 A
18 A
4 A
HV
380-480 V - 50/60 Hz
13 A
13 A
2.5 A