Technical data and dimensions
79
14.2.4 Electrical characteristics
For any change of power supply voltage in relation to the initial configuration, contact your service center.
*
UL 489 certified circuit breaker, supplied by the customer.
Circuit breaker with a minimum interrupting short circuit current
10 kA
GFI (or RCD) type B, differential circuit breaker compatible with type T.T
electrical network
1) For other networks, types T.N or I.T, use the proper protection device
100 mA
1)
Model
Main circuit breaker rat-
ing
*
(minimum values)
Cable wire specifica-
tions
A 124H
200-230V 50/60 Hz
15 A
2.5 mm
2
AWG-14
380-480V 50/60 Hz
10 A
1.5 mm
2
AWG-14
A 204H/A 204X
200-230V 50/60 Hz
25 A
6 mm
2
AWG-10
380-480V 50/60 Hz
15 A
2.5 mm
2
AWG-14
A 604H
200-230V 50/60 Hz
25 A
6 mm
2
AWG-10
380-480V 50/60 Hz
15 A
2.5 mm
2
AWG-14
A 804H/A 804X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 1204H
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 1504H/A 1504X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 1804H/A 1804X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 2404H/A 2404X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 2504H/A 2504X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
A 3004H/A 3004X
200-230V 50/60 Hz
32 A
6 mm
2
AWG-10
380-480V 50/60 Hz
25 A
6 mm
2
AWG-10
Summary of Contents for A 1204H