68
Malfunctions
W05/D05 LOG IN.1
W06/D06 LOG IN.2
W12/D12 LOG IN.3
Incorrect monitoring configuration
Check that [LOG 1], [LOG 2] or [LOG 3] is [DIS-
ABLED] if input is not used.
Check the warnings or fault delays.
Connection not present or cut
Check the sensors connection on LI1, LI2 and LI3
plug
Other problem
Contact your service center.
D07 LOW SPEED FB
Incorrect monitoring configuration
Check that [DEFINITION] [FREQUENCY IN-
VERTER] [FB Inverter] is set on [V1000] or [DES-
ABLED].
Wiring problem on
EMS
emergengy stop connec-
tor
Check that there is a strap between pins 2 and 3
of emergengy stop terminal
EMS
.
Rotor rotation is difficult or FB mechanical prob-
lem
Check the rotation of the FB shafts.
Free the FB manually
.
Frequency inverter trouble
Replace the frequency inverter, or
Contact your service center.
W09 LOW VOLTAGE
Main power is not correct
Check the main power.
Other problem
Contact your service center.
W10/D10 T° MOT FB
FB temperature is > 120°C
The cooling circuit is clogged
Check the water cooling supply on the pump.
Check that the motor is correctly cooled.
Check that the cooling pipes are not clogged.
Sensor or wiring faulty
Check the sensor connection on FB MOTOR plug
Problem on FB motor temperature sensors
Check the proper operation of FB motor tempera-
ture sensors (NC contacts).
Other problem
Contact your service center.
W11 INLET VALVE
Incorrect monitoring configuration
Check that [INLET VALVE] is [DISABLED] if no
valve is installed.
Faulty wiring
Check the valve connection on the
J4
connector
.
Malfunction
Check the proper operation opening/closing of the
valve.
Other problem
Contact your service center.
PW13/W13/PD13/D13 PURGE
FLOW
No purge flow
Check nitrogen supply.
Check the valve connection on N2 VALVES plug
Insufficient purge flow
Check the setting of nitrogen flow using pressure
regulator.
Check that there is no leak.
Incorrect threshold setting
Check the warning and alarm thresholds in [SET-
TING][PURGE][PURGE FLOW][LOW DE-
FAULTS].
Flowmeter not connected
Check the flowmeter connection on FLOW CONT
plug
Other problem
Contact your service center.
PW14/W14/D14 R1 P DRIVE
Roots 1 incorrect threshold setting
Check the warning and alarm thresholds in [SET-
TING][POWER][R1 POWER].
Roots 1 mechanical problem
Check the rotation of the Roots 1 rotors (if possi-
ble).
Problem with power reading
Check current and voltage on the monitoring and/
or at the frequency inverter output.
If the is no frequency inverter, check the current
sensor.
Problem on frequency inverter
Contact your service center.
Other problem
Contact your service center.
W15 /D15 HP FB T°
either
HP T° > setpoint T +5°C
FB HP water valve problem
Check the proper operation and the connection of
the FB HP water valve
.
Check the power supply of HP water valve coil on
FB.
Cooling problem
Check the water cooling supply on the pump.
Check that the cooling pipes and /or the cooling
plate are not clogged.
Temperature sensor and/or wiring faulty
Check the temperature reading on HHR as well as
the connections of the sensor on FB HP TEMP
plug and of the valve on FB VALVES plug
.
Other problem
Contact your service center.
Incident
Cause
Remedy