Malfunctions
67
11.2 Pump or Roots does not start
11.2.1 The white indicator light Power does not switch on
11.2.2 HHR display is not lighting up
11.3 The pump is running and HHR is showing a fault message
Incident
Cause
Remedy
Power
indicator light does not
switch on
No mains
Check power supply.
Main switch on position
O
Switch it on
I.
External emergency stop engaged
Check the presence of cover plug
EMS
or check
its locking.
Unit emergency stop engaged
Rotate to unlock the emergency stop on the front
panel.
Other problem
Contact your service center.
Incident
Cause
Remedy
HHR display is not lighting up, but
Power
indicator light is lit
Twisted cable incorrectly connected
Check the connection.
Other problem
Contact your service center.
Incident
Cause
Remedy
PW01/W01/PD01 EXH. PRESS
Exhaust pressure > Thresholds
Check the nitrogen flow rate setting value
Check that exhaust pipe line is not clogged.
Exhaust pressure < 550 hPa
Check the connection of the sensor on FB PRESS
plug
– Pressure sensor is faultly, replace it.
Exhaust pressure sensor faulty
Check the displayed pressure on HHR, if the read-
ing is inconsistent, replace the sensor.
Monitoring faulty
If the pressure sensor is operating but the dis-
played pressure is inconsistent, replace the moni-
toring.
Other problem
Contact your service center.
D01 EXH. PRESS
Exhaust pressure > 1990 hPa
or
The pump does not start and
the exhaust pressure > 1500 hPa
Check the nitrogen flow rate setting value
Check the exhaust pipe is not clogged.
Pressure sensor faulty
Check the displayed pressure on HHR, if the read-
ing is inconsistent, replace the sensor.
Monitoring faulty
If the pressure sensor is operating but the dis-
played pressure is inconsistent, replace the moni-
toring.
Other problem
Contact your service center.
PW03/W03/D03 FB P DRIVE
Incorrect FB threshold setting
Check the settings of power thresholds [SET-
TING][POWER][FB POWER].
FB mechanical problem
Check the rotation of the FB shafts
Free the FB manually.
Problem with power reading
Check current and voltage on the monitoring and/
or at the frequency inverter output.
If there is no frequency inverter, check the current
sensor.
Communication problem between monitoring and
FB frequency inverter
Depending on the trouble, replace the monitoring
or the frequency inverter.
Communication problem (pump without frequency
inverter)
Check the current sensor:
– the sensor is faulty, replace it.
Other problem
Contact your service center.
W04/D04 IN ANA1
W50/D50 USER ANA
Incorrect monitoring configuration
Inconsistent analog reading feedback
Check that [ANA 1], [ANA 2] or [ANA USER] is
[DISABLED] if input is not used.
Check the warning or fault thresholds.
Check that the voltage is in accordance with mon-
itoring display.
Sensor faulty
Check the proper operation of the sensors con-
nected to these analog inputs.
Other problem
Contact your service center.