Clarus 500 GC Installation Guide
99
7.
Connect the other end of the fused-silica tubing to the PID exit tubing as follows:
a)
Insert the nut on one end of the 1/16 to 1/16-in. union (P/N 0990-3075) over the
PID exit tube until it bottoms. Using two wrenches, tighten the nut 1/4 turn past
fingertight.
b)
Remove the nut and metal ferrule from the other end of the 1/16 to 1/16-in.
union. Put the metal ferrule aside; you will not need it.
c)
Insert the nut and a 1/16-inch graphite/Vespel ferrule (P/N 0992-0301) over the
free end of the fused-silica tubing.
d)
Insert the end of the fused-silica tubing into the union until it bottoms. Slide the
nut up to the union and tighten it fingertight.
e)
Using two 1/4-in. wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting.
Do not overtighten the nut!
You can operate the ElCD and the PID in series or operate either detector independently.
Always connect your analytical column to the PID.
Figure 31. Connecting the fused silica line to the ElCD and the PID.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......