Clarus 500 GC Installation Guide
101
Install the NPD Bead Assembly
If you purchased your Clarus 500 GC equipped with a Nitrogen Phosphorus Detector (NPD),
the NPD bead assembly is packaged separately and shipped inside the Clarus 500 GC oven.
To install the NPD bead assembly, use the following procedure and refer to Figure 32.
CAUTION
After installing a NPD bead, you must set up the NPD and condition a
new bead before you can use it. Refer to Setting Up a NPD in Chapter 8
of the Clarus 500 GC User’s Manual (P/N 0993-6073)
.
1.
Open the detector cover.
2.
Remove the bead assembly package from the Clarus 500 GC oven.
3.
Loosen the knurled ring on the NPD collector assembly. Then remove the collector
assembly.
CAUTION
You may find it easier to first remove the coaxial cable from the
collector assembly before you remove the collector assembly.
4.
Remove the screw that secures the bead transformer assembly to the top of the Clarus
500 GC oven.
5.
Plug the bead assembly (P/N N612-0092) into the connector on the bead transformer.
The connector is keyed so that the bead assembly can only be inserted one way.
6.
Carefully insert the bead portion of the bead assembly in the detector body as shown
in Figure 32.
7.
Secure the bead transformer to the top of the oven with the screw removed in step 4.
8.
Replace the collector assembly on the detector body and secure it by tightening the
knurled ring. If the coaxial cable was removed, connect it to the collector assembly.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......