Connecting the Accessories
98
Figure 30. Examples of a good tubing cut and bad cuts.
4.
Position the nut on the fused-silica tubing so that the back of the nut is
8.8 cm (3
1/2
inches) from the end of the fused-silica tubing. This positions the fused-
silica tubing inside the ElCD receiver.
5.
Using typewriter "white-out" or a felt-tipped pen, make a mark on the fused-silica
tubing just beyond the back edge of the nut.
CAUTION
To avoid contaminating the system, make sure the nut does not come
into contact with the mark on the fused-silica tubing.
6.
Connect the fused-silica tubing to the ElCD fitting inside the oven as follows:
a)
Insert this end of the fused-silica tubing into the ElCD fitting. Slide the nut up to
the ElCD and turn it until it is fingertight.
b)
Push the fused-silica tubing into the ElCD fitting until the mark on the tubing is
aligned with the back of the nut.
c)
Using two 1/4-inch wrenches, tighten the nut only until the fused-silica tubing
cannot be pulled out of the ElCD fitting.
Do not overtighten the nut!
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......