Preparing Your Laboratory
42
Gas Requirements:
All gases must have a minimum purity of 99.995%. Gases used with the mass spectrometer
and ELCD detectors require a minimum purity of 99.999%. Gas cylinders should be located
outside of the laboratory whenever possible and should always be stored and operated in the
vertical position. Always use copper tubing that is free of grease, oil, and organic material
for all gases delivered to the Clarus 500 GC.
Helium, Nitrogen,
8.5% H
2
/91.5% Helium
95% Argon/5% Methane:
A number 1A (200 ft
3
) gas cylinder should be used for
all carrier gases. Filter all gases (except methane)
through a moisture filter and/or hydrocarbon trap and
de-oxo filter. Argon/methane should be filtered
through a moisture filter and a de-oxo filter. Gas
delivery pressure to the GC should be 60 – 90 psig
(414-621 kPa).
Air:
A number 1A (200 ft
3
) gas cylinder of compressed air
or an air compressor can be used. All air should be
filtered through a moisture filter. Do
NOT
use
“Breathing Air." When using manual pneumatics, gas
delivery pressure to the GC must not exceed 30 psi
(207 kPa). If this is not possible, secondary regulation
will be required. With PPC pneumatics, a delivery
pressure range of 70 – 90 psig (483 - 621 kPa) is
acceptable.
Hydrogen:
A number 2 (62 ft
3
) gas cylinder or a hydrogen
generator can be used. All hydrogen should be filtered
through a moisture filter. When using manual
pneumatics, gas delivery pressure to the GC must not
exceed 30 psi (207 kPa). If this is not possible,
secondary regulation will be required. With PPC
pneumatics, a delivery pressure range of 60 – 90 (414-
621 kPa) psig is acceptable.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......