Connecting the Gases and Electrical Supply
80
Connecting the Gas for Subambient Operation
This section describes how to connect a liquid nitrogen or liquid carbon dioxide supply to
your Clarus 500 GC for subambient operation.
For liquid nitrogen, use a supply with a liquid delivery pressure of 20 – 30 psig. For liquid
carbon dioxide, use a size 1A cylinder equipped with a fitting containing a full-length dip
(eductor) tube and approximately 27.3 kg (60 lb) of carbon dioxide.
The LN
2
or CO
2
supply tubing connects the supply tank to the 1/4-in. union on the rear of the
Clarus 500 GC. To install the supply tubing, select the procedure that corresponds to the
subambient kit you are installing and refer to Figure 22.
Connect the LN
2
Supply Tubing
1.
Locate the LN
2
tank. Make sure the tank provides a liquid delivery pressure of 20 –
30 psig.
2.
Connect the large nut on one end of the LN
2
supply tubing (P/N N600-0403) to the
1/2-in. pipe to 1/4-in. Swagelok reducer labeled “LIQUID"
on the LN
2
tank by turning the nut clockwise until it is finger-tight. To provide a leak-
free connection, tighten the nut an additional half-turn with a large adjustable wrench.
3.
Connect the 1/4-in. nut on the other end of the tubing to the 1/4-in. union (see Figure
22) on the rear of the Clarus 500 GC by turning the nut clockwise until it is finger-
tight. To provide a leak-free connection, place a 9/16-in. wrench on the nut and a 1/2-
in. wrench on the union, then tighten the nut an additional half-turn.
4.
Turn on the valve and check for leaks.
Connect the CO
2
Supply Tubing
WARNING
Protect the CO
2
supply from undue heating. CO
2
is a gas above 31°C (87.8°F) at a
pressure of 1069 psig. The pressure increases rapidly at temperatures above 31°C
(87.8°F).
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......