Common Injector/Detector Gas Connections
Figures 13 through 21 shows how gases are connected to the Clarus 500 GC in several
common injector/detector configurations. Note that the strain relief, which is required on all
gas lines, is not shown in these figures.
WARNING
Position incoming gas lines away from the oven exhaust duct so that the hot
exhaust will not blow directly on the gas lines.
CAUTION
Thoroughly purge the incoming gas lines before measuring and adjusting the
gases. This is especially important when using hydrogen gas.
NOTE
:
For Clarus 500 GC’s that have a combination of both PPC and manual pneumatics, set the
detector gas delivery pressure between 60 and 90 psig and install the fixed 30-psig stainless
steel diaphragm regulator (P/N N610-1473), behind the Clarus 500 GC for the manual
pneumatics detector gases.
The manual pneumatics bulkhead always contains Swagelok nuts in all six locations whether
or not there is an injector or a detector connected to this bulkhead. The PPC modules will be
installed only if your Clarus 500 GC has been configured with PPC for an injector and/or
detector.
The Clarus 500 GC can be configured as any of the following:
•
All zones (injector/detector) as PPC.
•
All zones as manual pneumatics.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......