Clarus 500 GC Installation Guide
97
Make the ElCD/PID Series Connection
Use the following procedure to make the ElCD/PID series connection. After you make this
connection, you can operate either the ElCD or the PID without any loss of sensitivity.
NOTE
:
Always connect the analytical column to the PID, and the PID exit tube to the ElCD.
1.
Cut a length of the supplied deactivated 0.53-mm. I.D. fused silica tubing (P/N N930-
1358) long enough to connect the PID exit line to the ElCD. Try to use the shortest
possible length (approximately 12 inches).
2.
nsert a 1/16-in. column nut and a 1/16-in. graphite/Vespel ferrule
(P/N 0992-0301) over one end of the cut fused-silica tubing. Make sure the tapered
end of the ferrule faces the nut as shown below.
Figure 29. Placing a 1/16-in. nut and graphite ferrule over the fused silica line.
3.
Cut about 1 cm (3/8 inch) from the end of the fused-silica tubing using a wafer scribe
(P/N N930-1386, pkg. of 10 scribes) or other column cutting tool. Break off the tubing
at the score mark and make sure that the break is clean and square. Examine the cut
with a magnifying glass and compare it to the examples below.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......