Clarus 500 GC Installation Guide
103
1.
To connect a device to amplifier Channel A or B, first configure output 1 or 2 for a
recorder or integrator as described in Chapter 3, Initial Setup Procedures, in the
Clarus 500 GC User's Manual
.
2.
Open the small access cover on top of the Clarus 500 GC. With a large flat bladed
screw driver loosen the large screw. On the main cover is a screw well. Loosen this
screw with the large flat bladed screwdriver. You will then be able to lift the entire
cover up and have access to the recorder or integrator.
3.
To connect a recorder or integrator to Channel A, connect the ground lead of the
device to TB1-13 and the positive lead of the device to TB1-14. Make sure you route
the wires around the strain relief posts as shown in Figure 34.
4.
To connect a recorder or integrator to Channel B, connect the ground lead to TB1-15
and the positive lead to TB1-16.
Figure 33. Recorder and integrator connections at TB1.
In addition to the amplifier Channel A and B contacts, TB1 has READY and START
contacts. Table 3 describes the functions of these contacts.
Summary of Contents for clarus 500 gc
Page 1: ...Clarus 500 GC Installation Guide ...
Page 5: ...Introduction 1 ...
Page 6: ......
Page 12: ...Introduction 12 Label locations on instrument ...
Page 13: ...Safety Practices 2 ...
Page 14: ......
Page 33: ...Preparing Your Laboratory 3 ...
Page 34: ......
Page 43: ...Clarus 500 GC Installation Guide 43 Sample Preparation Requirements Customer Responsibility ...
Page 45: ...Installing the Clarus 500 GC 4 ...
Page 46: ......
Page 58: ...Installing the Clarus 500 GC 58 ...
Page 59: ...Connecting the Gases and Electrical Supply 5 ...
Page 60: ......
Page 82: ...Connecting the Gases and Electrical Supply 82 Figure 22 Connecting the CO2 or LN2 supply ...
Page 84: ...Connecting the Gases and Electrical Supply 84 Table 2 Plugs Used in Different Countries ...
Page 88: ...Connecting the Gases and Electrical Supply 88 ...
Page 89: ...Connecting the Accessories 6 ...
Page 90: ......
Page 92: ...Connecting the Accessories 92 Figure 24 ElCD Control Unit front panel ...
Page 106: ...Connecting the Accessories 106 ...
Page 107: ...PPC Restrictor Information 7 ...
Page 108: ......